BENEFIT FROM THE BLUETOOTH® CONNECTION

 

PIPETMAN® M CONNECTED

Chapter 10 

GUidELiNES FOr GOOd PiPETTiNG

Aspirate and Dispense

Fit new PIPETMAN DIAMOND Tip for the best results 

(refer to Chapter 19 

SPECiFiCATiONS 

on page 31).

Single Channel Model

For single channel models, push the tip holder into the tip 

using a slight twisting motion to ensure a firm and airtight 

seal.

Multichannel Model

For multichannel models, PIPETMAN DIAMOND Tips are 

best fitted with the ROCKY RACK technique, invented by 

Gilson, available only in our TIPACK and TOWERPACK. 

Pre-Wet the Tips 

All Models

Pre-wetting the tips before pipetting helps prepare the tips for the best pipetting 

performance. Ideally, the pre-wet includes both immersing the tip in the liquid and 

performing one pipetting step. Pre-wetting the tips helps ensure that volumes that 

you pipette will achieve accuracy and precision within specifications. 
Hold the pipette vertically and immerse 

the tip in the liquid (refer to Table 1). 

Press the push button to aspirate the set 

volume of liquid. Wait a few seconds (time 

depends on model, refer to Table 1); then 

withdraw the pipette tip from the liquid. 

You may wipe any droplets away from 

the outside of the tip using a medical 

wipe; however if you do so, take care to 

avoid touching the orifice of the tip.
Place the end of the tip against the inside 

wall of the recipient vessel (at an angle of 

10° to 40°). Press the push button. Wait 

for at least a few seconds before releasing 

the push button to expel any residual 

liquid from the tip. While removing the 

pipette draw the tip along the inside 

surface of the vessel.

 

 

MODEL 

IMMERSION 

DEPTH (mm)

WAIT TIME

(SECONDS)

P10M

1

1

P20M

2-3

1

P100M

2-4

1

P200M

2-4

1

P300M

2-4

1

P1200M

2-4

2-3

P5000M

3-6

4-5

P10mLM

5-7

4-5

P8x10M/P12x10M

1

1

P8x20M/P12x20M

2-3

1

P8x100M/P12x100M

2-4

1

P8x200M/P12x200M

2-3

1

P8x300M/P12x300M

2-3

1

P8x1200M/P12x1200M

2-4

2-3

Table 1

Immersion depth and wait time

 

Mold and cavity references are marked on the collar, ensuring the traceability. 

For quality assurance purposes batch numbers appear on all packages (bags and 

boxes).

 

They form an airtight seal with the tip holder, preventing the leaks that cause poor 

accuracy and a lack of precision.

 

PIPETMAN DIAMOND Tips (except filter tips) may be autoclaved at 121°C for  

20 minutes at 0.1 MPa.

NOTICE

do not autoclave PiPETMAN Filter Tips as the filter will be damaged.

NOTE

To ensure the best performance from your Gilson pipette, you should 

ALWAYS use PiPETMAN diAMONd Tips (in accordance with iSO8655) 

system, because PiPETMAN diAMONd Tips were used to establish the 

specifications.

Figure 7

Plastic adapter 

for D10 and 

DF10 tips

Figure 8

P8x10M and P12x10M

Long tips

Short tips

Ejector spacer

Ejector spacer D10

*A plastic adapter is required to eject 

D10 and DF10 tips (short tips). The 

adapter is supplied with P10M. No 

adapter is required when using DL10 

and DFL10 tips (long tips).

 

PIPETMAN M CONNECTED MULTICHANNEL

P8x10M

D10 

DL10

DF10ST 

DFL10ST

0.5–10 µL

0.5–10 µL

P12x10M

P8x20M

DL10

D200

DFL10ST

DF30ST

1–20 µL

1–20 µL

P12x20M

P8x100M

D200 

DF100ST

10–100 µL

5–100 µL

P12x100M

P8x200M

D200 

D300

DF100ST

DF200ST 

DF300ST

20–200 µL

5–200 µL

P12x200M

P8x300M

D200 

D300

DF200ST 

DF300ST

10–300 µL

10–300 µL

P12x300M

P8x1200M

D1200

DF1200ST

50–1200 µL

50–1200 µL

P12x1200M

 

PIPETMAN M CONNECTED SINGLE CHANNEL 

Model

PIPETMAN 

DIAMOND Tips

Gilson Specification

Volume Range

Standard 

PIPET Mode

REPETITIVE 

Mode

P10M

D10*

DL10*

DF10ST

DFL10ST

0.5–10 µL

0.5–10 µL

P20M

D200

DF30ST

2–20 µL

2–20 µL

P100M

D200

DF100ST

5–100 µL

5–100 µL

P200M

D200

D300

DF200ST

DF300ST

20–200 µL

5–200 µL

P300M

D200

D300

DF200ST

DF300ST

20–300 µL

10–300 µL

P1200M

D1000

D1200

DF1000ST

DF1200ST

100–1200 µL

20–1200 µL

P5000M

D5000

500–5000 µL 100–5000 µL

P10mLM

D10mL

1–10 mL

200 µL–10 mL

 

Slot

Summary of Contents for F81040

Page 1: ...User s Guide EN...

Page 2: ...D 18 Tip Ejector Extension for Use with P10M 18 Personal Label Name and or Application Tag 19 GOOD LABORATORY PRACTICE FEATURES 19 General 19 Cycle Counters 19 Alarms 20 POWER MANAGEMENT 20 Battery Ch...

Page 3: ...fatigue and increase pipetting efficiency PIPETMAN M Connected is a unique pipette with only two buttons to operate the pipette and reach all menu options PIPETMAN M Connected features the classic PIP...

Page 4: ...9 10 11 12 13 14 15 16 17 18 19 Upper part Push button and volume adjustment knob Menu button direct access to all menu functions Tip ejector button Display Elastomer plug to cover the battery chargi...

Page 5: ...the tip by pressing the ejection button Your PIPETMAN M Connected is ready for the next pipetting cycle while maintaining the volume and the last settings Continue pipetting as you would with your mec...

Page 6: ...g protocol first change the pipetting speed of your pipette refer to Chapter 8 PIPETTING SPEED CONTROL on page 15 Aspiration can be done by using the standard speed by default speed 6 Dispensing shoul...

Page 7: ...nables personalized pipetting protocol creation on a PC or tablet for transfer to PIPETMAN M Connected through PIPETMAN M Connected Software an easy to use software To use this mode on a PC the softwa...

Page 8: ...pending on your application You can change aspiration speed and dispensing speed independently from very slow to very fast speed 1 speed 6 Your speed selection is memorized for each pipetting mode unt...

Page 9: ...4 5 P10mLM 5 7 4 5 P8x10M P12x10M 1 1 P8x20M P12x20M 2 3 1 P8x100M P12x100M 2 4 1 P8x200M P12x200M 2 3 1 P8x300M P12x300M 2 3 1 P8x1200M P12x1200M 2 4 2 3 Table 1 Immersion depth and wait time Mold an...

Page 10: ...surface of the liquid refer to Immersion Depth and Wait Time Table 1 on page 17 Change the tip before aspirating a different liquid sample or reagent Change the tip if a droplet remains at the end of...

Page 11: ...thium Ion battery Operation on the battery is done at user s risk only Opening the pipette voids the warranty Alarms Low battery warning Service is overdue warning refer to Chapter 15 MAINTENANCE on p...

Page 12: ...mber of readjustments carried out by this pipette NOTE Changes of service settings will reset all the service info to zero Chapter 14 CONFIGURATION Enter the menu using the menu button You will find n...

Page 13: ...tion Enter the sub menu in the same way and select the ADJUSTMENT menu The adjustment menu allows you to access the following items Standard readjustment Reset settings Standard Readjustment This menu...

Page 14: ...Chapter 21 REPLACEMENT PARTS on page 34 To access the O ring remove the tip ejector and unscrew the lower part of the tip holder You should now be able to remove the O ring from the piston Sometimes t...

Page 15: ...er 16 CLEANING AND DECONTAMINATION PIPETMAN M Connected is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated As part of your quality system you...

Page 16: ...nents with distilled or sterile water 4 Leave the parts to dry by evaporation or wipe them with a clean lint free tissue or soft cloth 5 Lubricate the piston and reassemble the pipette according to th...

Page 17: ...e used with the recommended PIPETMAN DIAMOND Tips Under these conditions Gilson volumetric specifications in standard pipetting PIPET mode are guaranteed with a performance exceeding ISO 8655 2 recomm...

Page 18: ...22 F807022 F807022 F807022 F807022 F807022 F807022 Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006...

Page 19: ...ture and pressure for distilled water TEMPERATURE AIR PRESSURE HPA C 800 853 907 960 1013 1067 15 1 0018 1 0018 1 0019 1 0019 1 0020 1 0020 15 5 1 0018 1 0019 1 0019 1 0020 1 0020 1 0021 16 1 0019 1 0...

Page 20: ...R IMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL PUNITIVE INDIRECT OR IN...

Page 21: ...gilson com contactus LT801594 G 2021 Gilson Inc All rights reserved...

Reviews: