PIPETMAN® M CONNECTED 

|

 

USER’S GUIDE

TROUBLESHOOTING

 

PIPETMAN® M CONNECTED

Chapter 20 

SAFETY PrECAUTiONS ANd  

LiMiTATiONS OF USE

For safety reasons, it is important to observe the following instructions:

 

Battery and electrical specifications:

• 

Use AC adaptor and stand adapter indoors.

When pipetting infectious, radioactive, toxic and other hazardous solutions,

 please 

observe all the safety precautions (e.g. wear protective clothing, goggles and gloves) and 

regulations appropriate for your country.

 

Storage conditions 

 

Temperature: -20°C to 50°C - Humidity max: 80%

 

Temperature of use 

 

Between 4°C and 40°C (specifications may vary)

 

Use only genuine PiPETMAN diAMONd Tips

 and original Gilson accessories 

and spare parts. If the equipment is used in a manner not specified by Gilson in 

the user’s guide, the protection provided in the equipment may be impaired.

 

Equipment disposal 

 

This equipment must not be disposed of with unsorted municipal waste. Instead, it is 

your responsibility to correctly dispose of your waste equipment by handing it over to 

an authorized facility for separate collection and recycling. It is also your responsibility 

to decontaminate the equipment in case of biological, chemical, and/or radiological 

contamination so as to protect from health hazards the persons involved in the 

disposal and recycling of equipment. For more information about where you can drop 

off your waste equipment for recycling, please contact your local dealer from whom 

you originally purchased the product or your local council. By doing so, you will help 

conserve natural resources and you will ensure that your waste equipment is recycled 

in a manner that protects human health and the environment. Thank you.

 

PIPETMAN M Connected can be used indoor and outdoor,if precautions are respected 

as described above in this handbook and in the good laboratory practices (GLP).

NOTICE

We strongly recommend you fully charge the battery before using the 

pipette. PiPETMAN M Connected cannot be used until 50% of charge is 

reached. PiPETMAN M Connected is supplied with an AC adaptor that is 

suitable for your country. You must only use an original Gilson AC adaptor 

specific to this product. Charge the battery in the pipette, using the AC-

adapter or the stand adapter.

The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6):

 

Basis of adjustment, Ex.

 

Reference temperature, 20°C

 

Relative humidity, 50%

 

Barometric pressure, 101 kPa

 

Use of distilled water grade 3 (ISO 3696)

 

Ten measurements for each test 

volume, which are nominal volume, 

50% of nominal volume and 10% 

of nominal volume. (tested mode: 

PIPET, speed 6 with  

PIPETMAN DIAMOND Tips)

Table 2

PIPETMAN® M Connected maximum permissible errors 

WARNING

do not use PiPETMAN M Connected in a potentially explosive environment 

or with potentially explosive chemicals.

NOTICE

do not allow the liquid to enter the body of the pipette

 Gilson maximum permissible errors are guaranteed only when PIPETMAN pipettes are used with the recommended  

PIPETMAN DIAMOND Tips. 

Under these conditions, Gilson volumetric specifications in standard pipetting (PIPET mode) are guaranteed with a performance 

exceeding ISO 8655-2 recommendations for this mode. In the absence of ISO recommendations for repetitive pipetting mode for air 

displacement pipettes, Gilson volumetric specifications for repetitive pipetting (REPETITIVE mode) are guaranteed within ISO 8655-2 

recommendations for standard pipetting (Cf. ISO 8655-2 table 1).

CAUTION

PiPETMAN M Connected uses a Lithium-ion battery. Operation on the 

battery is done at user’s risk only. Opening the pipette voids the warranty. 

dispose of used batteries in accordance with legal regulations. Batteries 

may not be disposed of with household waste and may explode if 

disposed in fire.

 

Li-ion battery pack; 1 Ah / 3.6 V. Charging time: approx. 3 hours for a 

fully discharged battery (80% in one hour).

 

AC adaptor : input voltage, country-specific: 100-240 V, 50/60 hz 

(0.5A max.) Output voltage: +5 V dC, 3.5A (17.5W max.)

 

Class ii   this device is double insulated.

 

PIPETMAN M CONNECTED SINGLE CHANNEL

Model

PIPETMAN DIAMOND 

Tips

Part 

Number

Gilson Specifications

ISO 8655-2

Standard PIPET mode

REPETITIVE 

mode 

volume

Systematic  
error (µL)

Random 
 error (µL)

Volume Range

Volume 

(μL)

Systematic 
error (μL)

Random 
error (μL)

P10M

D10

DL10

DF10ST

DFL10ST

F81040

0.5–10 µL

0.5

1

5

10

± 0.040

± 0.025

± 0.060

± 0.080

 0.013

 0.012

 0.020

 0.025

0.5–10 µL

± 0.12

± 0.12

± 0.12

± 0.12

 0.08

 0.08

 0.08

 0.08

P20M

D200

DF30ST

F81041

2–20 µL

2

10

20

± 0.075

± 0.100

± 0.150

 0.025

 0.035

 0.050

2–20 µL

± 0.2

± 0.2

± 0.2

± 0.2

 0.1

 0.1

 0.1

 0.1

P100M

D200

D100T

F81042

5–100 µL

5

10

50

100

± 0.35

± 0.30

± 0.38

± 0.40

 0.10

 0.10

 0.12

 0.15

5–100 µL

± 0.8

± 0.8

± 0.8

± 0.8

 0.3

 0.3

 0.3

 0.3

P200M

D200

D300

DF200ST

DF300ST

F81043

20–200 µL

20

100

200

± 0.40

± 0.80

± 1.00

 0.15

 0.22

 0.26

5–200 µL

± 1.6

± 1.6

± 1.6

 0.6

 0.6

 0.6

P300M

D200

D300

DF200ST

DF300ST

F81044

20–300 µL

20

30

150

300

± 0.80

± 0.70

± 0.90

± 1.05

 0.16

 0.20

 0.23

 0.30

10–300 µL

± 4.0

± 4.0

± 4.0

± 4.0

 1.5

 1.5

 1.5

 1.5

P1200M

D1000

D1200

DF1000ST

DF1200T

F81045

100–1200 µL

100

120

600

1200

± 2.5

± 2.4

± 3.6

± 6.0

 0.4

 0.4

 0.8

 1.2

20–1200 µL

± 16

± 16

± 16

± 16

 6.0

 6.0

 6.0

 6.0

P5000M

D5000

F81046

500–5000 µL

500

2500

5000

± 10

± 15

± 25

 2

 4

 7

100–5000 µL

± 40

± 40

± 40

 15.0

 15.0

 15.0

P10mLM

D10mL

F81047

1–10 mL

1 mL 

5 mL

10 mL

± 25

± 30

± 50

 4

 8

 12

200 µL–10 mL

± 60

± 60

± 60

 30.0

 30.0

 30.0

 

 

PIPETMAN M CONNECTED MULTICHANNEL

P8x10M

D10 

DL10

DF10ST

DFL10ST

F81048

0.5–10 µL

0.5

1

5

10

± 0.05

± 0.04

± 0.08

± 0.10

 0.02

 0.02

 0.04

 0.06

0.5–10 µL

± 0.24

± 0.24

± 0.24

± 0.24

 0.16

 0.16

 0.16

 0.16

P12x10M

F81049

P8x20M

DL10

D200

DFL10ST

DF30T

F81050

1–20 µL

1

2

10

20

± 0.08

± 0.09

± 0.15

± 0.25

 0.05

 0.06

 0.10

 0.12

1–20 µL

± 0.4

± 0.4

± 0.4

± 0.4

 0.2

 0.2

 0.2

 0.2

P12x20M

F81051

P8x100M

D200

 

DF100ST

F81052

10–100 µL

10

50

100

± 0.25

± 0.50

± 0.80

 0.14

 0.20

 0.25

5–100 µL

± 1.6

± 1.6

± 1.6

 0.6

 0.6

 0.6

P12x100M

F81053

P8x200M

D200 

D300

DF100TS

DF200ST 

DF300ST

F81054

20–200 µL

20

100

200

± 0.50

± 1.00

± 2.00

 0.16

 0.30

 0.50

5–200 µL

± 3.2

± 3.2

± 3.2

 1.2

 1.2

 1.2

P12x200M

F81055

P8x300M

D200 

D300

DF200ST 

DF300ST

F81056

10–300 µL

10

30

150

300

± 1.00

± 1.00

± 1.50

± 2.40

 0.18

 0.18 

 0.375

 0.45

10–300 µL

± 8.0

± 8.0

± 8.0

± 8.0

 3.0

 3.0

 3.0

 3.0

P12x300M

F81057

P8x1200M

D1200

DF1200ST

F81058

50–1200 µL

50

120

600

1200

± 4.0

± 4.0

± 6.0

± 9.6

 0.7

 0.7

 1.5

 1.8

50–1200 µL

± 32

± 32

± 32

± 32

 12

 12

 12

 12

P12x1200M

F81059

Summary of Contents for F81040

Page 1: ...User s Guide EN...

Page 2: ...D 18 Tip Ejector Extension for Use with P10M 18 Personal Label Name and or Application Tag 19 GOOD LABORATORY PRACTICE FEATURES 19 General 19 Cycle Counters 19 Alarms 20 POWER MANAGEMENT 20 Battery Ch...

Page 3: ...fatigue and increase pipetting efficiency PIPETMAN M Connected is a unique pipette with only two buttons to operate the pipette and reach all menu options PIPETMAN M Connected features the classic PIP...

Page 4: ...9 10 11 12 13 14 15 16 17 18 19 Upper part Push button and volume adjustment knob Menu button direct access to all menu functions Tip ejector button Display Elastomer plug to cover the battery chargi...

Page 5: ...the tip by pressing the ejection button Your PIPETMAN M Connected is ready for the next pipetting cycle while maintaining the volume and the last settings Continue pipetting as you would with your mec...

Page 6: ...g protocol first change the pipetting speed of your pipette refer to Chapter 8 PIPETTING SPEED CONTROL on page 15 Aspiration can be done by using the standard speed by default speed 6 Dispensing shoul...

Page 7: ...nables personalized pipetting protocol creation on a PC or tablet for transfer to PIPETMAN M Connected through PIPETMAN M Connected Software an easy to use software To use this mode on a PC the softwa...

Page 8: ...pending on your application You can change aspiration speed and dispensing speed independently from very slow to very fast speed 1 speed 6 Your speed selection is memorized for each pipetting mode unt...

Page 9: ...4 5 P10mLM 5 7 4 5 P8x10M P12x10M 1 1 P8x20M P12x20M 2 3 1 P8x100M P12x100M 2 4 1 P8x200M P12x200M 2 3 1 P8x300M P12x300M 2 3 1 P8x1200M P12x1200M 2 4 2 3 Table 1 Immersion depth and wait time Mold an...

Page 10: ...surface of the liquid refer to Immersion Depth and Wait Time Table 1 on page 17 Change the tip before aspirating a different liquid sample or reagent Change the tip if a droplet remains at the end of...

Page 11: ...thium Ion battery Operation on the battery is done at user s risk only Opening the pipette voids the warranty Alarms Low battery warning Service is overdue warning refer to Chapter 15 MAINTENANCE on p...

Page 12: ...mber of readjustments carried out by this pipette NOTE Changes of service settings will reset all the service info to zero Chapter 14 CONFIGURATION Enter the menu using the menu button You will find n...

Page 13: ...tion Enter the sub menu in the same way and select the ADJUSTMENT menu The adjustment menu allows you to access the following items Standard readjustment Reset settings Standard Readjustment This menu...

Page 14: ...Chapter 21 REPLACEMENT PARTS on page 34 To access the O ring remove the tip ejector and unscrew the lower part of the tip holder You should now be able to remove the O ring from the piston Sometimes t...

Page 15: ...er 16 CLEANING AND DECONTAMINATION PIPETMAN M Connected is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated As part of your quality system you...

Page 16: ...nents with distilled or sterile water 4 Leave the parts to dry by evaporation or wipe them with a clean lint free tissue or soft cloth 5 Lubricate the piston and reassemble the pipette according to th...

Page 17: ...e used with the recommended PIPETMAN DIAMOND Tips Under these conditions Gilson volumetric specifications in standard pipetting PIPET mode are guaranteed with a performance exceeding ISO 8655 2 recomm...

Page 18: ...22 F807022 F807022 F807022 F807022 F807022 F807022 Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006...

Page 19: ...ture and pressure for distilled water TEMPERATURE AIR PRESSURE HPA C 800 853 907 960 1013 1067 15 1 0018 1 0018 1 0019 1 0019 1 0020 1 0020 15 5 1 0018 1 0019 1 0019 1 0020 1 0020 1 0021 16 1 0019 1 0...

Page 20: ...R IMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL PUNITIVE INDIRECT OR IN...

Page 21: ...gilson com contactus LT801594 G 2021 Gilson Inc All rights reserved...