LEAK TEST
|
PIPETMAN® M
The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6):
●
Basis of adjustment, Ex.
●
Reference temperature, 20°C
●
Relative humidity, 50%
●
Barometric pressure, 101 kPa
●
Use of distilled water grade 3 (ISO 3696)
●
Ten measurements for each test
volume, which are nominal volume,
50% of nominal volume and 10%
of nominal volume. (tested mode:
PIPET, speed 6 with
PIPETMAN DIAMOND Tips)
PIPETMAN M SINGLE CHANNEL — VARIABLE VOLUME MODELS
Model
PIPETMAN
DIAMOND Tips
Part
Number
Gilson Specifications
I
SO 8655-2
Standard PIPET mode
REPETITIVE
mode
volume
Systematic
error (µL)
Random
error (µL)
Volume Range
Volume
(μL)
Systematic
error (μL)
Random
error (μL)
P10M
D10
DL10
DF10
DFL10
F81022
0.5–10 µL
0.5
1
5
10
± 0.040
± 0.025
± 0.060
± 0.080
≤
0.013
≤
0.012
≤
0.020
≤
0.025
0.5–10 µL
± 0.12
± 0.12
± 0.12
± 0.12
≤
0.08
≤
0.08
≤
0.08
≤
0.08
P20M
D200
DF30
F81023
2–20 µL
2
10
20
± 0.075
± 0.100
± 0.150
≤
0.025
≤
0.035
≤
0.050
2–20 µL
± 0.2
± 0.2
± 0.2
± 0.2
≤
0.1
≤
0.1
≤
0.1
≤
0.1
P100M
D200
D100
F81013
5–100 µL
5
10
50
100
± 0.35
± 0.30
± 0.38
± 0.40
≤
0.10
≤
0.10
≤
0.12
≤
0.15
5–100 µL
± 0.8
± 0.8
± 0.8
± 0.8
≤
0.3
≤
0.3
≤
0.3
≤
0.3
P200M
D200
D300
DF200
DF300
F81024
20–200 µL
20
100
200
± 0.40
± 0.80
± 1.00
≤
0.15
≤
0.22
≤
0.26
5 – 200 µL
± 1.6
± 1.6
± 1.6
≤
0.6
≤
0.6
≤
0.6
P300M
D200
D300
DF200
DF300
F81014
20–300 µL
20
30
150
300
± 0.80
± 0.70
± 0.90
± 1.05
≤
0.16
≤
0.20
≤
0.23
≤
0.30
10–300 µL
± 4.0
± 4.0
± 4.0
± 4.0
≤
1.5
≤
1.5
≤
1.5
≤
1.5
P1200M
D1000
D1200
DF1000
DF1200
F81015
100–1200 µL
100
120
600
1200
± 2.5
± 2.4
± 3.6
± 6.0
≤
0.4
≤
0.4
≤
0.8
≤
1.2
20–1200 µL
± 16
± 16
± 16
± 16
≤
6.0
≤
6.0
≤
6.0
≤
6.0
P5000M
D5000
F81016
500–5000 µL
500
2500
5000
± 10
± 15
± 25
≤
2
≤
4
≤
7
100–5000 µL
± 40
± 40
± 40
≤
15.0
≤
15.0
≤
15.0
P10mLM
D10mL
F81017
1–10 mL
1 mL
5 mL
10 mL
± 25
± 30
± 50
≤
4
≤
8
≤
12
200 µL–10 mL
± 60
± 60
± 60
≤
30.0
≤
30.0
≤
30.0
PIPETMAN M MULTICHANNEL
P8x10M
D10
DL10
DF10
DFL10
F81025
0.5–10 µL
0.5
1
5
10
± 0.05
± 0.04
± 0.08
± 0.10
≤
0.02
≤
0.02
≤
0.04
≤
0.06
0.5–10 µL
± 0.24
± 0.24
± 0.24
± 0.24
≤
0.16
≤
0.16
≤
0.16
≤
0.16
P12x10M
F81026
P8x20M
DL10
D200
DFL10
DF30
F81027
1–20 µL
1
2
10
20
± 0.08
± 0.09
± 0.15
± 0.25
≤
0.05
≤
0.06
≤
0.10
≤
0.12
1–20 µL
± 0.4
± 0.4
± 0.4
± 0.4
≤
0.2
≤
0.2
≤
0.2
≤
0.2
P12x20M
F81028
P8x100M
D200
DF100
F81018
10–100 µL
10
50
100
± 0.25
± 0.50
± 0.80
≤
0.14
≤
0.20
≤
0.25
5–100 µL
± 1.6
± 1.6
± 1.6
≤
0.6
≤
0.6
≤
0.6
P12x100M
F81019
P8x200M
D200
D300
DF100
DF200
DF300
F81029
20–200 µL
20
100
200
± 0.50
± 1.00
± 2.00
≤
0.16
≤
0.30
≤
0.50
5–200 µL
± 3.2
± 3.2
± 3.2
≤
1.2
≤
1.2
≤
1.2
P12x200M
F81030
P8x300M
D200
D300
DF200
DF300
F81031
10–300 µL
10
30
150
300
± 1.00
± 1.00
± 1.50
± 2.40
≤
0.18
≤
0.18
≤
0.375
≤
0.45
10–300 µL
± 8.0
± 8.0
± 8.0
± 8.0
≤
3.0
≤
3.0
≤
3.0
≤
3.0
P12x300M
F81032
P8x1200M
D1200
DF1200
F81020
50–1200 µL
50
120
600
1200
± 4.0
± 4.0
± 6.0
± 9.6
≤
0.7
≤
0.7
≤
1.5
≤
1.8
50–1200 µL
± 32
± 32
± 32
± 32
≤
12
≤
12
≤
12
≤
12
P12x1200M
F81021
Table 2
PIPETMAN® M maximum permissible errors
Gilson maximum permissible errors are guaranteed only when PIPETMAN pipettes are used with the recommended
PIPETMAN DIAMOND Tips.
Under these conditions:
●
Gilson volumetric specifications in standard pipetting (PIPET mode) are guaranteed with a performance exceeding
ISO 8655-2 recommendations for this mode.
●
In the absence of ISO recommendations for repetitive pipetting mode for air-displacement pipettes, Gilson volumetric
specifications for repetitive pipetting (REPETITIVE mode) are guaranteed within ISO 8655-2 recommendations for standard
pipetting (Cf. ISO 8655-2 table 2 above).