P
O
W
ER
M
A
N
A
G
E
M
EN
T
19
PIPETMAN® M
|
USER’S GUIDE
Working with Charger Connected to the Pipette
PIPETMAN M can also be used for pipetting while charging.
1.
Plug the power supply into your pipette.
2. Press the push button to start your pipette.
PIPETMAN M is ready to pipette.
Low Battery Warning
PIPETMAN M has a low battery alert. As the battery runs down, the
low battery message appears. The warning is replaced by a blinking
symbol of an empty battery. If it is not recharged, the pipette will switch off after a
while. Settings will not be lost.
If the battery is completely discharged (black screen, no operation) connect the
pipette to the charger for at least five minutes, and then a “batt. too low” message
will appear. The pipette will display the last settings used a few minutes later and will
continue charging.
Changing the Battery
If the following symptoms occur, please contact your Gilson service center:
●
PIPETMAN M battery won’t charge or will not operate even if connected to the
charger at least for 10 minutes.
●
PIPETMAN M battery provides you very short cycle time or permanently needs
to be used in connected mode.
NOTE
You can also recharge PIPETMAN M by USB with your PC; however it will
take much longer than by power supply.
NOTICE
Use only the original power transformer supplied by Gilson. Use of an
incompatible power supply can damage your PIPETMAN M.
Do not operate PIPETMAN while connected to a PC.
CAUTION
PIPETMAN M uses a Lithium-Ion battery. Operation on the battery is done
at user’s risk only. Opening the pipette voids the warranty.