Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
MetPak/MetPak RG and MetPak Pro Base Stations
Page | 69
Issue 1
Doc. No. 1723-PS-0023
October 2018
aMC3!
Temperature (degrees
C), Humidity (%),
Dewpoint (Degrees C)
and CRC.
atttn<CR><LF>
00053<CR><LF>
0D0!
Retrieve a line of the
above data.
a<temp><humidity><dewpoint>
<status><CRC><CR><LF>
0+022.2+042.7+009.0+00LV{<CR><L
F>
aMC4!
Pressure (Hecto
Pascals/millibars) and
CRC.
atttn<CR><LF>
00051<CR><LF>
0D0!
Retrieve a line of the
above data.
a<pressure><CRC><CR>
<LF>
0+1017.8GM <CR><LF>
aMC5!
PRT, Analogue Input 1,
Analogue Input 2,
Digital Input 1 and
CRC.
atttn<>CR><LF>
00054<CR><LF>
0D0!
Retrieve a line of the
above data.
a<prt><AI1><AI2><dig1><
CRC><CR>< LF
0-001.19+0000.6+0000.4+0000.000
CQx <CR><LF>
aMC6!
PRT and CRC.
atttn<CR><LF>
00051<CR><LF>
0D0!
Retrieve a line of the
above data.
a<prt><CRC><CR><LF>
0-001.19IT@<CR><LF>
aMC7!
Analogue Input 1,
Analogue Input 2 and
CRC.
atttn<>CR><LF>
00052<CR><LF>
0D0!
Retrieve a line of the
above data.
a<AI1><AI2><CRC><CR>
<LF
0+0000.6+0000.4FE~<CR><LF>
aMC8!
Digital Input 1 and
CRC.
atttn<CR><LF>
00051<CR><LF>
0D0!
Retrieve a line of the
above data.
a<dig1><CRC><CR><LF>
0+0000.000EgO<CR><LF>
aMC9!
Power Voltage (Volts)
and CRC
atttn<CR><LF>
00011<CR><LF>
0D0!
Retrieve a line of the
above data.
a<voltage><CRC><CR>
<LF>
0+12.2K@R<CR><LF>