40
10.1.2 Calibrating the column in wireless mode
The column may be calibrated in wireless mode
Proceed as follows
Attention:
During calibration the column must not be loaded. Calibrate only when there is
not equipment mounted on the column!
Attention:
During calibration the columns lift range will be driven through and set. The
column will drive slowly to the bottom position and then to the top position and
finally return to the starting position.
Before starting, ensure that the column’s
lift rage is in no way obstructed!
Attention:
The Dolly will only function when the HCU and electronic unit are paired. If
either unit is swapped or exchanged the units must be newly paired!
Switch on the electronic unit by pressing the ON/OFF button on the housing.
As soon as the electronic is activated (LED is blinking red/green), press and hold the SPD-
button on the HCU.
Then switch the column on by pressing the ON/OFF button on the HCU whilst holding the
SPD button pressed.
As soon as the HCU LED is on the SPD-button can be released.
The
Display shows „CAL DOLLY“.
Info:
To stop the calibration procedure press either EXIT or the ON/OFF button on
the HCU.
Attention:
When the HCU is turned off it automatically turns of the electronic unit too. To
turn back on follow the described procedure again i.e. first turn on the
electronic unit, then the HCU..
Press the CAL button (RMP) button to begin calibrating.
Attention:
During calibration the columns lift range will be driven through and set. The
column will drive slowly to the bottom position and then to the top position and
finally return to the starting position.
Before starting, ensure that the column’s
lift rage is in no way obstructed!
During calibration the column will drive slowly to the bottom position and then to the top position and
finally return to the starting position and the display will show “PLEASE WAIT”. When the display
shows “DONE” press the EXIT button to leave the calibration mode.
Summary of Contents for GF-Primo Dolly
Page 18: ...17 GF Primo s 3 steering modes and selector position...
Page 20: ...19 GF Secondo s three steering modes and their selector positions...
Page 27: ...26 Handle Rosette Battery cover Exchangeable HCU cover HCU battery Battery cover lock...
Page 34: ...33 Key combination to change the channel Change the channel...
Page 37: ...36...
Page 42: ...41 Tastenkombination zum Kalibrieren des Dolly...
Page 60: ...59 MARKERs 3 to 6 can be selected in the same manner Setting MARKERs...
Page 68: ...67 Recording a sequence...
Page 70: ...69...
Page 88: ...87...