74
o
the HCU battery is in place. The wireless mode does not work without battery.
o
both the HCU and electronic are sending and receiving on the same channel.
Before operating the column and the HCU for the first time it is necessary to synchronize both units.
Cable mode:
Check if
o
the emergency OFF Switch is inserted into the off position. If so, twist clockwise to
deactivate.
o
the cable is connected correctly.
o
the connectors are damaged (broken or bent pin).
o
the same problem occurs with a different cable.
Error:
Column and HCU don’t operate in cable mode.
Solution:
Check if
o
the emergency OFF Switch is inserted into the off position. If so, twist clockwise to
deactivate.
o
the cable is connected correctly.
o
the connectors are damaged (broken or bent pin).
o
the same problem occurs with a different cable.
o
the battery poles on the column are extended and not flush with column.
o
the batteries are charged.
Error:
The column cannot be calibrated.
Solution:
Check if
o
the correct button combination is being pressed to activate the calibration mode (se
e „4.2
“calibrating the column” on 10).
o
the emergency OFF Switch is inserted into the off position. If so, twist clockwise to
deactivate.
o
the batteries are charged.
Check the operational mode!
Wireless mode:
Check if
o
the electronic is switched on. In wireless mode it must be switched on first.
o
t
he HCU is equipped for wireless operation and not that it’s only equipped for cable
operation.
o
the HCU battery is in place. The wireless mode does not work without battery.
o
both the HCU and electronic were synchronized (se
e „4.1 Synchronizing the HCU and
Column on page 8 )and that they are sending and receiving on the same channel.
If the above solutions do not result in an improvement change to cable mode and repeat the
calibration procedure
.
Cable mode:
Check if
o
the cable is connected correctly.
o
the connectors are damaged (broken or bent pin).
o
the same problem occurs with a different cable.
Summary of Contents for GF-Primo Dolly
Page 18: ...17 GF Primo s 3 steering modes and selector position...
Page 20: ...19 GF Secondo s three steering modes and their selector positions...
Page 27: ...26 Handle Rosette Battery cover Exchangeable HCU cover HCU battery Battery cover lock...
Page 34: ...33 Key combination to change the channel Change the channel...
Page 37: ...36...
Page 42: ...41 Tastenkombination zum Kalibrieren des Dolly...
Page 60: ...59 MARKERs 3 to 6 can be selected in the same manner Setting MARKERs...
Page 68: ...67 Recording a sequence...
Page 70: ...69...
Page 88: ...87...