![Geokon GK-604D User Manual Download Page 98](http://html.mh-extra.com/html/geokon/gk-604d/gk-604d_user-manual_2214579098.webp)
86
CA = M
×
RINT
×
SA
CB = M
×
RINT
×
SB
DA = (CA
×
cos(ZZ))-(CB
×
sin(ZZ))
DB = (CA
×
sin(ZZ))+(CB
×
cos(ZZ))
Equation E-4 Profile Calculation
Note: Accumulate (
Σ
) DA and DB results at each depth increment (from the bottom
up or the top down) to obtain the profile.
E.3. GTILT Users
When using GTILT with the GK-604D use a Probe Constant of 10000 for both
English and Metric probes when using 2.0sin Units! For 2.5sin Units use a Probe
Constant of 12500.