45
NOTE: An “Auto Record” survey can be paused at any time and re-
started as long as the probe is moved to the proper level reflected
by the “Level:” display. When an Auto Record survey is paused,
the data can still be recorded in the “normal” fashion by tapping
on the “Record Data” buttons.
3.3.5.6 Finish Survey with:
This parameter deals exclusively with “unfinished” survey data files. A
survey is unfinished if readings were not taken at each level from starting
up to the zero level. Many surveys may not completely finished because
the geometry of the probe will not allow the last reading or two to be taken
while the probe is still in the casing.
Valid choices for this selection (see Figure 40) include:
Nothing The survey will not be filled in and will remain unfinished.
NaN(s) Each missing level “row” of the survey will be filled in with
“NaN(s)” which is the floating point representation of a non-
numerical value. NaN is an abbreviation for ”Not a Number”.
:
:
:
#READINGS:11
FLEVEL, A+, A-, B+, B-
5.0, 45, 12, -87, 81
4.5, 46, 12, -84, 81
4.0, 44, 12, -85, 82
3.5, 46, 13, -85, 82
3.0, 44, 13, -86, 82
:
:
:
#READINGS:11
FLEVEL, A+, A-, B+, B-
5.0, 32, 25, -81, 82
4.5, 33, 24, -81, 82
4.0, 30, 24, -81, 82
3.5, 30, 24, -80, 81
3.0, 29, 24, -80, 81
2.5, 29, 24, -81, 81
2.0, 30, 24, -83, 81
1.5, NaN, NaN, NaN, NaN
1.0, NaN, NaN, NaN, NaN
0.5, NaN, NaN, NaN, NaN
0.0, NaN, NaN, NaN, NaN
Summary of Contents for GK-604D
Page 2: ......
Page 4: ......
Page 12: ......
Page 16: ...4 Figure 4 6000 6100 type probe Figure 5 GK 604 4 Interface ...
Page 28: ...16 Figure 11 Windows Mobile Device Center ...
Page 46: ...34 Figure 35 Viewing Compass Data ...
Page 52: ...40 Figure 41 Stable Indication Figure 42 Unstable Indication ...