91
APPENDIX G. Tiltmeter Operation (Model 6101)
When connected to the GK-604-4 Probe Interface Module (see Figure 5), the Model
6101 Tiltmeter (see Figure 78) can be read with the FPC-1 using the GK-604D IRA.
Figure 78 - Model 6101 Tiltmeter with 6201-3 Interface Cable
The Model 6101 Portable Tiltmeter is designed to be placed on an alignment plate
(Tiltplates, 6201-1X, see Figure 79) that has been permanently attached to the structure
being monitored. Measurements can be made on horizontal or vertical surfaces. The
readings are taken in pairs, 180 degrees apart from each other, to eliminate any
instrument bias and thereby obtain true tilt.