![Geokon GK-604D User Manual Download Page 104](http://html.mh-extra.com/html/geokon/gk-604d/gk-604d_user-manual_2214579104.webp)
92
Figure 79 - Tiltplates: 6201-1C (ceramic), 6201-1A (Copper plated Aluminum), 6201-1S (stainless)
It is assumed that a valid Bluetooth pairing exists between the GK-604-4 Interface
Module and the FPC-1 (see section 2.2 for more information about establishing
communication with the Interface Module). The recommended steps for connecting to
and taking a reading with the Model 6101 Tiltmeter are as follows:
1.
Connect one end of the 6201-3 cable to the Tiltmeter
2.
Connect the other end of the 6201-3 cable to the GK-604-4 Interface Module.
3.
On the FPC-1, launch the GK-604D IRA and create a new probe configuration:
•
Using the Context Menu (see section 3.2.1), after highlighting the Probe
Library, select the “Add Probe” menu item to create a new probe.
•
Name the new probe and select “Tiltmeter” for probe type.
•
Using the calibration sheet as a guide, enter the “Zero Shift A” and “Gage
Factor A” parameters. Leave the B channel parameters at zero.
•
Tap “Save Settings” to save the new probe configuration. See section 4.2 for
more information about probe configuration.