This procedure assumes that the board is installed in an operating system, and that the
±10V input range is selected.
1.
Connect the digital multimeter between the VTEST OUTPUT (+) and VTEST
RETURN (-) pins in the system I/O connector. Refer to Table 2.2.2 for pin
assignments.
2.
If power has been removed from the board, apply power now and wait at least 15
minutes before proceeding.
3.
Select the ±10V input range.
4.
Verify that the digital multimeter indication is +9.9000 VDC ±0.0009 VDC. If the
indication is not within this range, adjust the REFERENCE ADJUSTMENT trimmer
until the digital multimeter indication is within the specified range.