TS4000H
67
5.9.31 MOD1 Hardware Total Receive Errors (0x002A)
A read indicates the total Hardware UART errors that occurred in the slave device. The
maximum count is 65535 and then the counter will rollover to zero and begin counting again.
The total errors are an accumulation of the individual communication errors of registers 0x0027
– 0x0029.
5.9.32 MOD1 Hardware Clear Errors (0x002C)
A write 0 resets all MOD1 Hardware UART communication errors of registers 0x0027 – 0x002A
to 0. Only a write of value “0” is allowed for this register.
5.9.33 MOD1 Clear Communication Errors (0x002D)
A write 0 resets all MOD1 Total Data Errors, Function Code Errors, Starting Address Errors,
CRC High Byte Errors, and CRC Low Byte Errors to 0. Only a write of value “0” is allowed for
this register.
5.9.34 Clear Interface Module Communication Errors (0x002E)
A read indicates the total number of Interface Module to Base Unit Communication Errors.
The maximum count is 65535. Beyond 65535 the count resets to zero and begins counting
again. A write resets this value to 0. Only a write of value “0” is allowed for this register.
5.9.35 MOD2 Total Receive Errors (0x0043)
A read indicates the total Modbus Communication Receive Errors that occurred in the slave
device. The maximum count is 65535. Beyond 65535 the count resets to zero and begins
counting again.
5.9.36 MOD2 Total Data Errors (0x0044)
A read indicates the total number of illegal data write attempts that occurred in the slave device.
The maximum count is 65535. Beyond 65535 the count resets to zero and begins counting
again.
5.9.37 MOD2 Function Code Errors (0x0045)
A read indicates the number of Function Code Errors that occurred in the slave device.
The maximum count is 65535. Beyond 65535 the count resets to zero and begins counting
again.
5.9.38 MOD2 Starting Address Errors (0x0046)
The counter is incremented for each address that does not equal the device address.
A read indicates the number of Starting Address Errors that occurred in the slave device.
The maximum count is 65535. Beyond 65535 the count resets to zero and begins counting
again.