TS4000H
65
5.9.20 Sensor Scale (0x0018)
A read returns the current sensor scale. Sensor scale value is determined automatically by
sensor type, and cannot be written over Modbus. For a list of available sensor scales and
associated sensor types, refer to Table 34
Sensor Type
Sensor ID
Full Scale Value
0
No Sensor Detected
None
1
O
2
25% v/v
2
CO – 100
100 ppm
3
CO – 500
500 ppm
4 Cl
2
– 10
10 ppm
5
ClO
2
3 ppm
6 HCl 20
ppm
7
NO
100 ppm
8 NO
2
20
ppm
9
NH
3
50
50 ppm
10
NH
3
100
100 ppm
11
Reserved
N/A
12 SO
2
-20 20
ppm
13 SO
2
-100 100
ppm
14
H
2
S – 20
20 ppm
15
H
2
S-100
100 ppm
16
H
2
- 500
500 ppm
17-18
Reserved
N/A
19
Cl
2
– 20
20 ppm
20
H
2
S – 50
50 ppm
Table 34: Sensor Scale
5.9.21 Sensor Type (0x0019)
A read returns the current sensor type. A write is allowed only if the current sensor is set to:
“0” (no sensor)
“9” or “10” (any SOOO2
“2” or “3” (any CO sensor)
“4” or “19” (any CL
2
sensor)
“12” or “13” (any NH
3
sensor)
“9” or “10” (any NH
3
sensor)
If the sensor type is set to:
“0”, only sensor type value of “1” can be written, defining O
2
sensor
“2” or “3”, only sensor type values of “2” and “3” can be written, defining CO – 100
ppm and CO – 500 ppm scale sensors, respectively
In all other cases, the sensor type is determined automatically when the sensor is plugged in.
Write attempts in such cases will return a data Exception Code 03 (Illegal Data Value).