Moxi V
™
User Guide
Page 58
Test Error: Fluid was not
Detected
Sample not loaded or volume too
small
Make sure sample is loaded and volume is at
least 60 µL.
Test Error: Incomplete Test
Insufficient fluid (<60 µL)
Adjust pipette to ensure sufficient volume for the
test. For improperly calibrated pipettes, this
might require a setting >60 µL
Cells/beads too large
Make certain that the particle type is within the
specified size range of the cassettes being used
Highly/strongly aggregated cells
clogging filter and blocking flow
Try breaking apart cells using pipette trituration
and/or protease treatment.
Test Error: Stop Detected before
Start
Used cassette or cassette error.
Try the other side of the cassette or a new
cassette
Used or Invalid cassette
Previously used sensor detected
Do not reuse sensors.
User/System Defaults loaded
New Firmware has been loaded
No problem, simply touch OK to acknowledge
Instrument has detected corruption
or new version of firmware and
reset all settings to factory defaults
Contact ORFLO technical support
Warning: Low Battery
Battery needs to be recharged
Recharge instrument overnight while powered on
and plugged into the wall outlet. After that, make
sure to run future tests with the instrument
plugged into the AC adapter.
Warning: High Concentration
Particle count exceeds cassette
recommended concentration range
Dilute sample (e.g. 10x) and re-run
Warning - Time/Date lost
Battery ran out of power and
system clock reset
Charge the unit overnight and reset the
Date/Time in the
Settings
section (icon off of
Home
screen)