Moxi V
™
User Guide
Page 15
appending to an existing batch run). The selected option is display in orange. The other
options in green.
Cell QC – Single Run
By default, the system sets the “File Option” as a “Single Run” (single test, see image
above/middle). Below the file option selection, there are three test parameters that the user can
set: 1.) “File Name” 2.) “Sample ID” and 3.) “Dilution Factor”. The system defaults the test to
have a name of “QC-XXX” where XXX is a sequential number starting at “000” and incrementing
with each test run. The system will auto-enumerate that number with the lowest available
number (not already used by a saved test on the disk). This value also provides the test file
name (with an .fcs extension, e.g. “QC-004.fcs”). This file name field can be changed by
touching in the box and using the keyboard below to type a new name.
Notes: 1.) Capital
letters can be typed by hitting the “CAPS” key to change the keyboard to capital letters. 2.)
Additional punctuation/symbols are available by touching “Punc”.
The “Sample ID” field can be
used as general text description of the sample being run. This field can be changed by touching
in the box and using the keyboard below to type a description. The “Dilution Factor” field is a
drop-down list with the options of entering “1:1” (no dilution), “1:2” (2x dilution), “1:5” (5x
dilution), and “1:10” (10x dilution).
Note: the value selected here will automatically be
applied to the test count output to adjust the measured counts for the dilution level,
providing the true (dilution adjusted) initial starting concentration.
The value is set by
touching the box (image below/right) and touching one of the dropdown entries. The “Enter”
button can then be touched to start the test.