V 10/17
38
•
Troubleshooting
MagicCompact BA04
Problem/error/malfunction
Procedures/remedy
The pressure increase is indicat-
ed on the filter pressure gauge
Pressure increase on the filter
cartridges
Switch off the gun control units,
wait until the differential pressure
returns to normal again. Check
the cleaning cycles by ear, if
necessary, shorten the pause
times in the cleaning cycle.
Check, if the cleaning pressure is
set on 5 bar on the pressure in-
put valve (see also the section
"Troubleshooting" in the After Fil-
ter operating manual)
ATTENTION:
If the pressure gauge shows a
pressure rise higher than 3
kPa, contact your Gema agent
immediately!
Alarm has been released:
Message
Fan overpressure
Minimum pressure in filter hous-
ing not reached – corresponding
pressure gauge responding
Too little pressure, too much ex-
haust air, because too little or no
air resistance
Filter housing door is open
Sieve machine not fitted tightly
on the cyclone separator
Waste container not fitted tightly
(see also the troubleshooting
section in the After Filter manual)
Alarm has been released:
Message
Fan low pressure
Maximum pressure in the filter
housing exceeded – correspond-
ing pressure gauge responding
Pressure too high, insufficient
exhaust air because the air re-
sistance is too high
Filter clogged (valves defect or
cleaning pressure too low, at
least 5 bar)
Poor compressed air quality
(contains oil or water)
Malfunctions on running-in, until
the filter cake is built up on the
filter cartridges (see also the
"Troubleshooting" section in the
After Filter operating manual)
Alarm has been released:
Message
Guns not OK
Diagnostic adaptor of the guns
indicates, that no high voltage is
being produced
Turn on the gun control unit, or
correct the fault in the gun control
unit or gun with the information in
the corresponding operating in-
structions