GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
Chapter 10 - Care and Maintenance
10-17
P R E L I M I N A R Y
10-5-4-2
General Probe Cleaning Instructions
•
Acetone
•
Ammonium chloride
•
Alcohol, ethanol, isopropanol, or methanol
•
Bleach
•
Detergent
•
Hydrogen peroxide
•
Iodine
•
Para-hydroxybenzoic acid
•
Aloe vera, perfume, lanolin, or mineral oil
Do not soak or wipe the probe lens with any product listed above (see Warning). Doing so could result
in irreparable damage to the probe. Follow the care instructions supplied with each probe. In order to
disinfect or sterilize probes without damaging them:
•
Use gas sterilization and legally marked, sterile, pyrogen-free probe sheaths on probes used for
surgery.
•
Use cold chemical sterilization suitable for plastic medical scanning devices.
•
Do not immerse a probe beyond the ridge on its case.
•
Avoid cleaning and storage temperatures over 60
°
C (140
°
F).
WARNING
WARNING
Do
not
clean or immerse probes using any of the following products:
WARNING
WARNING
Neurological procedures must NOT be performed on a patient with Creutzfield-Jacob
disease because probes that have been used for such examinations cannot be
adequately sterilized.
WARNING
WARNING
Neurological and intraoperative procedures must be performed with the use of legally
marketed, sterile, pyrogen-free probe sheaths. Probes used during neurological
surgery must NOT be sterilized with liquid chemical sterilants because of the possibility
of neurotoxic residue remaining on the probe.
Summary of Contents for Vivid S5 N
Page 2: ......
Page 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Page 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Page 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Page 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Page 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Page 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Page 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Page 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Page 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Page 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Page 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Page 833: ......