GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
8-128
Section 8-6 - Lower Section Components Replacement Procedures
P R E L I M I N A R Y
8-6-4-6-1
Patient Archive Transfer Procedure
1.) If the original hard disk is still working, perform the backup procedure and save all presets to a
storage device.
2.) Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
.
3.) Disconnect the AC cable and battery.
4.) Remove the currently-installed hard disk from the Vivid S5 N / Vivid S6 N
system as described in
“Hard Disk Removal Procedure” on page 8-122
5.) Reconnect the AC cable and battery and power ON the Vivid S5 N / Vivid S6 N ultrasound unit.
6.) Perform a full software installation procedure, as described in
“Software Loading” on page 8-207
.
7.) Install new passwords provided by OTR to enable all options purchased by customer.
Follow the instructions for
“Software Options Configuration” on page 3-95
.
8.) Restore Connectivity Settings (refer to
“Connectivity Settings” on page 8-133
For step-by-step instructions, see
“Connectivity Configuration” on page 3-109
9.) Shut down the system.
10.)Disconnect all peripherals (USB and PCMCIA) from the system.
11.)Plug a Technical Support plug (dongle) into the USB port and boot-up the system. Exit to Windows,
as described in .
12.)Connect the original hard disk (the hard disk that you previously removed from the system) to the
appropriate port of the SATA/PATA adaptor. See the example in
CAUTION
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO
TAKING THE NECESSARY ESD PRECAUTIONS:
1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE
ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE
SCANNER (TO THE RIGHT OF THE POWER CONNECTOR).
2.
FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC
SENSITIVE EQUIPMENT.
Figure 8-158 PATA Hard Disk Connected to the USB-to-SATA/PATA Adaptor
Activity LED
Summary of Contents for Vivid S5 N
Page 2: ......
Page 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Page 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Page 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Page 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Page 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Page 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Page 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Page 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Page 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Page 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Page 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Page 833: ......