Chapter 3 - System Setup
3-127
P R E L I M I N A R Y
3.) Select the
Gateway Modality Worklist
service.
4.) Click the
Check
button.
-
If connectivity was achieved, then a green check should be displayed next to the name of the
Gateway's Modality Worklist service.
-
If connectivity was not achieved, then there will be a red X next to the name of the Gateway's
Modality Worklist service.
Troubleshoot Connectivity with the Gateway
If connection between the Vivid S5 N or Vivid S6 N and the HIS failed, the following are typical causes:
1.) No network connection.
2.) IP address and/or port number was incorrectly entered when configuring the Gateway Modality
Worklist service.
3.) Mitra support personnel have not yet configured the Gateway to accept Modality Worklist queries
from the Vivid S5 N or Vivid S6 N.
Verify that the Vivid S5 N or Vivid S6 Ncomputer name matches the value specified for the Gateway
Modality Worklist service and that it also matches the value supplied to the Mitra support personnel.
Configure the Export to HL7 Path
1.) Enter the configuration screen for the EchoPAC PC.
2.) Press the
Connectivity
button.
3.) Select the
Tools
tab.
4.) In the section labeled
Export Paths
, make the following changes:
a.) For the Export To HL7 path, enter the following information:
\\ < Name or IP Address of the Vivid HL7 Gateway>\POLL_DIR
Example: \\Gateway\POLL_DIR
b.) Verify that Text is select for the Export file format.
Test Connectivity with the Vivid HL7 Gateway
1.) Start a new study and acquire an image, or use an existing study.
2.) In the Measurements mode, make some of the measurements.
3.) In the Report screen, create and store a report.
4.) In the Patient screen, from the Examination List, select the study in which you created the report.
5.) From the
More
pull-down menu, select
HL7 Results
.
If the export was successful, the following message is displayed:
Export exam status: OK.
If the export was unsuccessful, the following message (or similar) is displayed:
Export exam status: Unable to copy file.
Summary of Contents for Vivid S5 N
Page 2: ......
Page 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Page 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Page 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Page 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Page 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Page 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Page 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Page 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Page 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Page 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Page 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Page 833: ......