GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4-27
P R E L I M I N A R Y
1
Trackball to
TR32 BF Digital Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
3
Under the
Special Instructions
area, activate the
Loops
checkbox and select the required number of
Loops
from
the combo box.
4
Adjacent to
On Fail,
select either
Stop
or
Continue
, as
required.
5
Trackball to the
Start
button and press
Set
The test commences.
When the TR32 BF Digital Test sequence is complete,
the View Test Log window opens automatically, listing
the test performed and showing whether it passed or
failed - see the examples in
and
6
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
System BF Digital Test
1
Trackball to
System BF Digital Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
3
Under the
Special Instructions
area, activate the
Loops
checkbox and select the required number of
Loops
from
the combo box.
4
Adjacent to
On Fail,
select either
Stop
or
Continue
, as
required.
5
Trackball to the
Start
button and press
Set
The test commences.
When the System BF Digital Test sequence is
complete, the View Test Log window opens
automatically, listing the test performed and showing
whether it passed or failed - see the examples in
6
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
ADC Test
1
Trackball to
ADC Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
3
Under the
Special Instructions
area, activate the
Loops
checkbox and select the required number of
Loops
from
the combo box.
Table 4-12 Front End Tests (cont’d)
Step
Task
Expected Result(s)
Summary of Contents for Vivid S5 N
Page 2: ......
Page 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Page 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Page 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Page 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Page 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Page 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Page 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Page 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Page 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Page 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Page 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Page 833: ......