D
IRECTION
5813707-100, R
EVISION
2
V
ENUE
™ S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-5
P R E L I M I N A R Y
8-2-3
Back Cover Replacement Procedure
8-2-3-1
Tools
Phillips screwdriver of appropriate size.
8-2-3-2
Time Required
5 min
8-2-3-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in
Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-3-4
Back Cover Removal Procedure
1) Remove batteries.
2) Remove Probe Holder Rubber by gently pulling it upwards starting from the middle.
FRU Part #
Refer to
.
•
Figure 8-73 Probe Holder Rubber Removal