Chapter 8 - Replacement Procedures
P R E L I M I N A R Y
8-2-24
ECG USB Cable Replacement Procedure
8-2-24-1
Tools
Phillips screwdriver of appropriate size
8-2-24-2
Time Required
5 min
8-2-24-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
8-2-24-4
ECG USB Cable
Removal Procedure
4.) Remove the system from the cart.
5.) Disconnect USB cable from the ECG assembly.
FRU Part #
Refer to
.
•
Figure 8-140 Disconnect ECG USB cable