D
IRECTION
5813707-100, R
EVISION
2
V
ENUE
G
O
™ S
ERVICE
M
ANUAL
Chapter 3 - System Setup
3-47
P R E L I M I N A R Y
3) Tap to select the DICOM storage item and to display the DICOM storage properties.
Figure 3-29 DICOM Storage Properties Window
4) Check the option
Allow SR
to enable DICOM SR.
The following additional options are available:
•
Allow SR private data
: send the current exam data in a private format. This option is by default
unchecked and should only be used with DICOM storage devices that can handle private data
format.
•
No images
: no images are sent, only M&A.
•
Signed Doppler velocities
: send signed Doppler velocities.
•
Use older SR version
: when checked a
Use older SR version
pull-down menu is displayed.
The current exam data will be sent in the same format as the selected SR version. Details about
format and content of the SR version can be found in the corresponding user manual of the
selected version.
These settings apply to both “TID 5200 Echocardiography Procedure Report” and “TID 5100
Vascular Ultrasound Procedure Report”
5) Tap
Apply
.
Summary of Contents for Venue Go