D
IRECTION
5813707-100, R
EVISION
2
H
AWAII
™ S
ERVICE
M
ANUAL
5-20
Chapter 5 - Venue Go™Components and Function (Theory)
P R E L I M I N A R Y
Section 5-10
Cooling System
5-10-1
General Information
The Venue Go™ system has 2 fans for system cooling:
To allow efficient cooling, the Scanner cooling consists of the following elements:
•
Heat Pipe to transfer the heat (installed on the C-BEB).
•
Outlet Blower at the edge of the Heat Pipe.
•
Fan to push cold air into scanner (installed on the C-FEB Heat Sink).
All fans are controlled and have two variable speeds.
The cooling requirement for the Venue Go™ ultrasound scanner with on board peripherals, is up to 600
BTU/h. This figure does not include cooling needed for lights, people, or other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/h demand on the cooling system.