TMO2D Digital Display User’s Manual
27
Chapter 4. Basic Programming
4.4.6 Set Display?
Although this feature is not commonly used, the TMO2D, when used with the TMO2, can display compensation values
for pressure or the percentage of background gas. For the analyzer to display these values, the TMO2 must include
external compensation. In addition, you must enable either background or pressure compensation through the
Calibration
menu (see page 37) in the
Select Compensation
submenu (page 38). If neither pressure nor background
compensation is enabled, a “Comp not enabled” message will appear.
4.4.7 Set Communications?
Baud Rate: The display contains a bi-directional, industry-standard RS-232C serial port that enables users to operate
the instrument remotely with a keypad or computer that supports the RS-232C protocol. All keypad operations, and
most display operations, can be performed remotely through this port. (See page 20 for operation instructions.)
The display supports the communication rates of 9600, 4800, 2400, 1200 and 300 baud. The default setting is 9600
baud. Other parameters are fixed at 8 bits, 1 stop bit, no parity.
Note:
The 300-baud setting is provided for compatibility with older equipment; however, the use of 300 baud greatly
limits the computation speed of the TMO2D, and its use is not recommended.
The
Set Communications?
submenu enables you to select both the baud rate and the update rate, when they
communicate with the TMO2D through the RS-232 port.
SETUP MENU
Set Display?
Press [YES] to set the display.
Display Compensation
[[NO]NE] pres bkgd
Use the [NO] or arrow keys to move the brackets to the desired entry. Press [YES] to
confirm the entry and exit.
SETUP MENU
Set Display?
Press [NO] to proceed to the next submenu.
SETUP MENU
Set Communications?
Press [YES] to enter the submenu.
SET COMMUNICATIONS
Set Baud Rate?
Press [YES] to enter the baud rate.
SELECT BAUD RATE
9600 baud?
Use the [NO] or arrow keys to scroll through the choices.
SELECT BAUD RATE
4800 baud?
SELECT BAUD RATE
2400 baud?
SELECT BAUD RATE
1200 baud?
Summary of Contents for TMO2D
Page 1: ...GE Measurement Control Solutions Gas Moisture TMO2D Digital Display User s Manual ...
Page 4: ...ii ...
Page 10: ...Chapter 1 Features and Capabilities 2 TMO2D Digital Display User s Manual ...
Page 24: ...Chapter 2 Installation 16 TMO2D Digital Display User s Manual ...
Page 26: ...Chapter 2 Installation TMO2D Digital Display User s Manual 18 ...
Page 72: ...Chapter 6 Specifications 64 TMO2D Digital Display User s Manual ...
Page 74: ...Appendix A Outline and Installation Drawings 66 TMO2D Digital Display User s Manual ...
Page 86: ...Appendix B Menu Maps 78 TMO2D Digital Display User s Manual ...
Page 92: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 84 ...
Page 96: ...Index 88 TMO2D Digital Display User s Manual ...
Page 98: ...Warranty 90 TMO2D Digital Display User s Manual ...
Page 100: ......
Page 101: ......