TMO2D Digital Display User’s Manual
41
Chapter 4. Basic Programming
4.8.3 Gas Calibration? (cont.)
Note:
Percent gas concentrations MUST be in ascending order, with Point #1 the lowest concentration.
The display shows “Working . . .”
The display will then show “Working . . .” before returning to the
Gas Calibration
submenu.
4.8.4 Pressure Calibration?
The
Pressure Calibration?
submenu enables users to directly calibrate the pressure transducer. While this menu option
is always displayed, it should only be programmed and used when the TMO2D is used with a TMO2 calibrated for
atmospheric pressure compensation.
Repeat this sequence for the remaining points. After you have entered the data, the display will show “Working . . .”
and return to the start of the submenu.
X.XX% O
2
, Curve #1
Enter Point? [YES] no
The display will show the data entered for Curve #1, and ask if you wish to enter a
point. Use the [NO] or arrow keys to scroll between choices, and press [YES] to
confirm either a [YES] or NO selection. (If you select [NO], the menu proceeds to
the next point entered.)
X.XX % O
2,
Curve #X
O2 mA [X.XX]
Use the numeric keys to enter the corresponding O
2
input milliamp value for the
given point, then press [YES] twice to confirm the entry.
X.XX % O
2,
Curve #X
Bkgd mA [X.XX]:
Use the numeric keys to enter the corresponding background gas compensation
input milliamp value for the given point. Press [YES] twice to confirm the entry, and
repeat the procedure for every point in the background grid.
CALIBRATION MENU
Gas Calibration?
Press [NO] to proceed to the
Pressure Calibration
submenu.
CALIBRATION MENU
Pressure Calib?
Press [YES] to enter the “Pressure Calibration?” submenu.
PRES XDUCER CURVE ENTRY
# of Points [X]:
Use the numeric keys to enter the number of points, and press [YES] twice to
confirm the entry.
Point # 1
mmHg [XXX.X]:
Use the numeric keys to enter the pressure in mmHg for the given point, and press
[YES] twice to confirm the entry.
Point # 1
Pres mA [X.XX]:
Use the numeric keys to enter the corresponding mA value for the point, and press
[YES] twice.
CALIBRATION MENU
Pressure Calib?
Press [NO] to proceed to the Auto Cal submenu.
Summary of Contents for TMO2D
Page 1: ...GE Measurement Control Solutions Gas Moisture TMO2D Digital Display User s Manual ...
Page 4: ...ii ...
Page 10: ...Chapter 1 Features and Capabilities 2 TMO2D Digital Display User s Manual ...
Page 24: ...Chapter 2 Installation 16 TMO2D Digital Display User s Manual ...
Page 26: ...Chapter 2 Installation TMO2D Digital Display User s Manual 18 ...
Page 72: ...Chapter 6 Specifications 64 TMO2D Digital Display User s Manual ...
Page 74: ...Appendix A Outline and Installation Drawings 66 TMO2D Digital Display User s Manual ...
Page 86: ...Appendix B Menu Maps 78 TMO2D Digital Display User s Manual ...
Page 92: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 84 ...
Page 96: ...Index 88 TMO2D Digital Display User s Manual ...
Page 98: ...Warranty 90 TMO2D Digital Display User s Manual ...
Page 100: ......
Page 101: ......