TMO2D Digital Display User’s Manual
33
Chapter 4. Basic Programming
4.6.2 Alarm Type (cont.)
The calibration error could occur if the TMO2D performs an autocal on a transmitter and the amount of drift exceeds a
programmed limit. You can set the limits in
Change AutoCal Limits?
on page 59.
The Cal occurred alarm trips if autoverification is enabled and an automatic calibration is performed on a transmitter. If
the amount of drift exceeds the limits programmed, the TMO2D stores the new drift calibration data and trips an alarm
to indicate the change. For further details, refer to page 43 and
Change AutoCal Limits?
on page 59.
Trip on High%?
[NO] yes
Use the [NO] or arrow keys to indicate whether or not you wish the alarm to trip on
the high percentage, and press [YES] to confirm the selection. (Selecting [NO]
returns the program to the Set High Alarm Function? entry.)
Alarm X, HIGH Setpoint
%O
2
[XX.X]:
Use the numeric keys to enter the desired value. Then press [YES] to enter the data,
and [YES] again to confirm the entry. (The X’s represent the previously entered
setpoint.)
ALARM FEATURES
Set High Alarm Function?
Press [NO] to proceed to the system fault action entry.
ALARM FEATURES
Set System Fault Action?
Press [YES] to enter the system fault action function. Signal faults occur if the signal
from the transmitter drops below 1 mA or exceeds 24 mA.
Trip On System Faults?
[NO] yes
Use the [NO] or arrow keys to indicate whether or not you wish the alarm to trip on
system faults, and press [YES] to confirm the selection.
ALARM FEATURES
Set System Fault Action?
Press [NO] to proceed to the cal error action entry.
ALARM FEATURES
Set Cal Error Action?
Press [YES] to enter the cal error action function.
Trip On Auto Cal Errors?
[NO] yes
Use the [NO] or arrow keys to indicate whether or not you wish the alarm to trip on
autocalibration errors, and press [YES] to confirm the selection.
ALARM FEATURES
Set Cal Error Action?
Press [NO] to proceed to the cal occurred action entry.
ALARM FEATURES
Set Cal Occurred Action?
Press [YES] to enter the cal occurred action function.
Trip On New Cal Data?
[NO] yes
Use the [NO] or arrow keys to indicate whether or not you wish the alarm to trip on
new calibration data, and press [YES] to confirm the selection.
ALARM FEATURES
Set Cal Occurred Action?
Press [NO] to proceed to the normal/failsafe mode function.
Summary of Contents for TMO2D
Page 1: ...GE Measurement Control Solutions Gas Moisture TMO2D Digital Display User s Manual ...
Page 4: ...ii ...
Page 10: ...Chapter 1 Features and Capabilities 2 TMO2D Digital Display User s Manual ...
Page 24: ...Chapter 2 Installation 16 TMO2D Digital Display User s Manual ...
Page 26: ...Chapter 2 Installation TMO2D Digital Display User s Manual 18 ...
Page 72: ...Chapter 6 Specifications 64 TMO2D Digital Display User s Manual ...
Page 74: ...Appendix A Outline and Installation Drawings 66 TMO2D Digital Display User s Manual ...
Page 86: ...Appendix B Menu Maps 78 TMO2D Digital Display User s Manual ...
Page 92: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 84 ...
Page 96: ...Index 88 TMO2D Digital Display User s Manual ...
Page 98: ...Warranty 90 TMO2D Digital Display User s Manual ...
Page 100: ......
Page 101: ......