7. Specifications
Page 178
DMS 2 Operating Manual
7.2.3
Recommended probes for the D-Multi Mode
Model
Designation
Nominal
Contact
Temperature
Measuring Range
Frequency
Diameter
Range
in Steel
HT400A/400
High Temperature
5 MHz
.500”
<1000°F
.100”-2.000”
(12.7mm)
(<538°C)
(2.5-50mm)
KBA560-D
General Purpose
5 MHz
.625”
<450°F
.100”-2.000”
(15.9mm)
(<230°C)
(2.5-50mm)
DA301
General Purpose
5 MHz
.475”
<140°F
.070”-1.800”
(12.1mm)
(<60°C)
(1.8-46mm)
DA312
Thin Materials
10 MHz
.300”
<130°F
.030”-.900”
(7.6mm)
(<54
°
C)
(.8-23mm)
DP-104
High Penetration
1 MHz
1.25”
<130°F
.100”-.2000”
(31.8mm)
(<54°C)
(2.5-50mm)
DM-401-GP
General Purpose
5 MHz
.700”
<300°F
.060”-8.00”
(17.8mm)
(<150°C)
(1.5-200mm)
Note:
Other probes listed in the D-Multi Mode probe parameter in the TG Setup Menu are compatible
over a limited thickness range.
NOTE:
Probe specifications are subject to change without notice.