4-12
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Maintenance:
Disassembly Guidelines
Display/Keyboard Assembly
Removal
1. Remove the top cover following the procedures above.
2. Disconnect the three cables connecting the display/keyboard
assembly to the main PCB.
NOTE
Two of these cables have locked connectors that must be lifted up
to release the cables.
3. Working from the outside of the top, remove the two TORX mounting
screws located on the right side of the assembly.
4. Remove the two TORX screws from the hinge bracket.
5. Remove the screw from the display ground at the left of the hinge
rod.
6. Slide the display hinge (metal rod) to the left to release it from the
mounting detent. A flat blade screw driver may be used to help slide
the rod.
7. Slightly lift up on the right hand side of the display/keyboard
assembly, and pull the assembly to the right to free the tabs from
their mounting slots. Do not lift the right side of the display too high
or the plastic tabs may be damaged.
8. When free from the main unit, the display/keyboard assembly can be
separated in to two pieces allowing replacement of either the
keyboard or display assembly.
NOTE
Further disassembly of the LCD assembly is not recommended.
Replace as complete assembly.
Two TORX
Mounting
Screws
Hinge
Tabs
43A
Roll Pin
Two TORX
Hinge
Bracket
Screws
Display Ground
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......