Installation:
MAC 5000 ST Requirements and Configuration
MAC 5000 ST Requirements and Configuration
Following is a list of interface requirements and setup configurations
required for the devices listed when used with the MAC 5000 ST option.
Compatible Blood Pressure Units
Colin - Model ST-780
Connection Requirements - Use cable PN 2008112-001 to connect from
the MAC5000 port
1
to the Colin serial port. Use cable PN 2008111-001
to connect from the MAC 5000
ANA/TTL
port to the Colin QRS trigger
input.
Device Configuration Requirements - None
MAC 5000 Configuration Requirements – At the
Main Menu
complete
the following in the order shown below:
Select
System Setup
,
Enter
System password
,
Exercise
Test
,
Inputs/Outputs
,
Change
Blood Pressure
to
Nipon-Colin
.
Sun Tech - Model Tango
Connection Requirements - Use cable PN 2008113-001 to connect
from the MAC 5000 port
1
to the Sun Tech serial port. Use cable PN
2008111-001 to connect from the MAC5000
ANA/TTL
port to the
Sun Tech QRS trigger input.
Device Configuration Requirements – At the Tango Main Menu complete
the following in the order shown below:
Select
Utilities
,
Select
Device
,
Scroll to
ECG Trigger
and press enter,
Scroll to
DIGITAL
↑
and press enter,
Scroll to
EXIT
and press
Enter
,
Scroll to
Test Parameters
and press
Enter
,
With
Technique
highlighted, press
Enter
,
Scroll to
DKA
and press
Enter
,
Scroll to
EXIT
and press
Enter
,
Scroll to
EXIT
and press
Enter
to return to the display screen.
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......