![background image](http://html.mh-extra.com/html/ge/mac-5000/mac-5000_service-manual_128654126.webp)
5-24
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Troubleshooting:
Input and Output Connectors
Input and Output Connectors
The following pages detail the input/output signals for those connectors.
The pin-by-pin descriptions identify the signal names and pin outs for
each connector on the unit.
A Pins (J1)
COM1 (COM3/4) Pins (J3)
Table 2. A Pins (J1)
Pin
Name
1
Data
2
NC
3
Ground
4
+5V
5
Clock
6
NC
Table 3. COM1 (COM3/4) Pins (J3)
Pin
COM1 Signal
COM3/4 Signal
1
RTS
COM3 TxD
2
CTS
COM3 RxD
3
TxD
4
Ground
5
RxD
6
DTR
COM4 TxD
7
+12V
8
DSR
COM4 RxD
2
1
6
5
3
4
66A
6
3
1
8
5
4
2
7
67A
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......