4-18
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Maintenance:
Disassembly Guidelines
Service Only Setups
1.
From the
Main Menu
, select
System Setup
.
2.
Press
Shift
+
F2
at the
System Setup
menu.
3.
Enter the service password and press
Enter
.
4.
The
Service Only Setup
window is displayed.
5.
Enter the serial number of the system.
NOTE
This is the number which was used when the option codes for this system
were generated. The number entered here must match the serial number on
the label of the system.
6.
Select/verify that
No
is selected for
Update Primary Boot
.
7.
Enter the
Print head resistance
. This number can be found on the print head
label.
8.
Select the appropriate language in the
Keyboard
menu.
9. Select
Return
.
Restore System Setups
1.
Power up the cart.
2.
From the
Main Menu
, select
System Setup
.
3.
Enter the
System Password
and press
Enter
.
4.
Select
Restore Setup
from the
System Setup
menu.
5.
Insert the floppy disk with the system setups which was set aside in before you
removed the previous CPU board.
6.
Select
From Diskette
from the
Restore Setup
menu.
Service Only Setup
Serial number:
_ _ _ _ _ _ _ _ _ _ _ _ _
Update Primary Boot:
_ _ _
Print head resistance:
_ _ _
Keyboard:
_ _ _ _ _ _ _ _ _
Return
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......