4-22
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Maintenance:
Disassembly Guidelines
Trolley Casters
Removal
1. Remove the MAC 5000 and all loose items from the trolley and then
place the trolley on its side.
2. Locate the slot under the arrow on the bearing dust cap and using a
small blade screwdriver, pry the cap from the caster to be removed.
3. Using an Allen wrench, remove the wheel shaft and wheel from the
caster.
4. Using an Allen wrench, remove the bolt holding the caster to the
trolley.
54A, 55A
56A
57A
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......