Revision B
MAC
™
5000 resting ECG analysis system
B-3
2024917-010
Appendix B – Technical Specifications:
Technical Specifications
Technical Specifications
Display
Item
Description
Type
264mm (10.4 in.) diagonal graphics backlit AM LCD
Resolution
640 x 480 pixels, with waveform enhancement
Displayed Data
Heart rate, patient name, ID, clock, waveforms, lead labels, speed, gain and filter settings,
warning messages, prompts and help messages
Computerized Electrocardiograph
Item
Description
Instrument Type
15 lead (14 channel) microprocessor-augmented, automatic electrocardiograph
Analysis Frequency
500 samples/s (sps)
ECG Storage
150 ECG’s (typical), 200 (maximum) on removable media (1.44 MB, 3.5” diskette)
Digital Sampling Rate
4000 samples/s/channel
Analysis
Pediatric and vectorcardiography
Optional: 12SL analysis, HI-RES and PHI-RES late potential analysis
Pre-acquisition
Provides 10 s of instantaneous ECG acquisition
Dynamic Range
AC differential:
±5 mV
DC offset:
±320 mV
Resolution
4.88 µV/LSB @ 250 sps, 1.22 µV/LSB @ 500 sps
Frequency Response
–3 dB @ 0.01 to 150 Hz
Common Mode Rejection
>140 dB (123 dB with AC filter disabled)
Input Impedance
>10 M
Ω
@ 10Hz, defibrillator protected
Patient Leakage Current
<10 µA
Pace Detection
750 µV @ 50 µs duration, Orthogonal LA, LL and V6
Special Acquisition Functions
Disconnected lead detection, electrode impedance, AC noise, baseline wander, and
muscle tremor
Communication
MAC and MUSE system compatible RS-232
Optional: Modem, FAX, and wireless transmission, remote retrieval (remote query)
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......