4-14
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Maintenance:
Disassembly Guidelines
Main CPU Board
Removal of CPU Board
NOTE
Before you begin, save the current System Setups to a diskette and
print System Setup report. This will be used to restore the system
setups after replacement of the CPU board.
If the system cannot be powered on, you may want to save the system
setups from another MAC 5000 that has identical system setups.
1. Remove the battery.
2. Remove the top cover assembly following the procedures in
3. Remove the display/keyboard assembly following the procedures in
4. Disconnect all remaining cable connections to the main PCB.
These include cables to the following:
power supply
printhead
battery connect PCB
floppy drive
acquisition module cable
5. With a TORX driver, remove the mounting screws holding the main
PCB in place. They are located around the outside edges of the main
PCB. Set screws aside for mounting new board.
6. Unlock the floppy cable connector by lifting the connector lock.
7. Remove the harness cable.
8. Lift the main PCB from the unit.
Reassembly of CPU Board
1. Insert the new CPU board in place and mount using the screws set
aside during disassembly.
2. Reassemble the top cover and display/keyboard assemblies by
reversing the steps for removal.
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......