Revision B
MAC
™
5000 resting ECG analysis system
5-23
2024917-010
Troubleshooting:
Frequently Asked Questions
Entering Patient Data
Q: Do I have to enter all the Information I see on the Patient Data
screen?
A: In
System Setup
>
Basic System
>
Patient Questions
you can require
that the patient identification number, or medical record number) be
entered. It is not a requirement to enter any other data. However, we
recommend that you enter the patient name and identification
number, at the least. If you are transmitting to the MUSE system
you will want to enter the
Location
number as well. If an emergency
situation dictates that you must complete the test. without entering
the patient data, make sure you edit the record to add the missing
information before you transmit it to the MUSE system.
Transmission
Losing Fields When Transmitting
Q: Why do I lose the Referring MD and Technician names off of my
reports when I transmit records to the MUSE system?
A: Your MAC 5000 may be transmitting to the SDLC modem on the
MUSE system instead of the CSI modem. Check in
System Setup
to
make sure you are transmitting to the MUSE system CSI phone
number.
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......