Revision B
MAC
™
5000 resting ECG analysis system
B-5
2024917-010
Appendix B – Technical Specifications:
Technical Specifications
Hi-Res and PHi-Res Signal-Averaged Electrocardiography
Item
Description
Frequency Response/Input
-3 dB @ 0.01 and 250 Hz
Frequency Response/Output
Upper Limit
Lower Limit
250 Hz
0.01, 25, 40 or 80 Hz
Sensitivities
Raw Data and Template
Average Beat
Filtered Signals and Vector
Magnitude
20 mm/mV
20 mm/mV and 50 mm/mV
1 mm/µV
Analysis Sampling Rate
1000 samples per second per channel
Digital Sampling Rate
4000 samples per second per channel
High/Low Pass Filters
Spectral filter using Fast Fourier Transform (FFT)
ADC Resolution
1.22 µV/LSB
Analysis Resolution
0.1525 µV/LSB
Physical
1
Item
Description
Height
9.4 cm (3.7 in) with display closed
Width
38.1 cm (15.0 in)
Depth
35.1 cm (13.8 in)
Weight
6.8 kg (15 lb) without paper
1. without trolley
Environmental
Item
Description
Operating Conditions
Temperature
Relative Humidity
Atmosphere Pressure
10° C to 40° C (50° F to 104° F)
1
20% to 95% RH noncondensing
700 to 1060 hPa
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......