Revision B
MAC
™
5000 resting ECG analysis system
5-21
2024917-010
Troubleshooting:
Frequently Asked Questions
Battery Capacity
Q: What is the capacity of the battery?
A: We recommend that the MAC 5000 be plugged into a wall outlet
whenever it is not in use. However, the life of the battery is
approximately 100 ECGs and one-page reports or six hours of
continuous operation (without printing).
System Setup
Location Number
Q: When entering in the patient data, how do I get the Location field to
automatically populate with the same number?
A: The
Location
number can be set in
System Setup
to save you from
entering it for each test.
Go to
System Setup
.
Select
Basic System
.
Select
Miscellaneous Setup
.
Arrow down to
Location
and type in the number you want set as
your default. Then press
Enter
.
Press the
esc
key until you return to
System Setup
.
Select
Save Setup
.
Select
To System
.
You can select
Main Menu
to exit
System Setup
.
Patient Questions
Q: How do I change what questions I see when I am entering the patient
data?
A: The patient questions you see on the
Patient Data
window when
starting a test were set up in
System Setup
.
Go to
System Setup
.
Select
Basic System
.
Select
Patient Questions
.
Select the patient questions you want to include when entering
the patient data for a test.
Press the
esc
key until you return to
System Setup
.
Select
Save Setup
.
Select
To System
.
You can select
Main Menu
to exit
System Setup
.
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......