![GE LOGIQ S7 Expert Service Manual Download Page 91](http://html.mh-extra.com/html/ge/logiq-s7-expert/logiq-s7-expert_service-manual_128037091.webp)
GE H
EALTHCARE
DRAFT
LOGIQ™ S7 E
XPERT
/P
RO
D
IRECTION
5460683
, R
EVISION
3
DRAFT (J
UNE
25, 2014)
S
ERVICE
M
ANUAL
Chapter 3 - Setup Instructions
3-29
3-7-4-2
Connecting the CW Pencil Probe
Insert the probe connector into the probe port all the way seated in. Carefully position the probe cord so
it is free to move and is not resting on the floor.
3-7-4-3
Cable Handling
Take the following precautions with probe cables:
•
Keep free from wheels
•
Do not bend the cable acutely
•
Avoid crossing cables between probes.
3-7-4-4
Activating the Probe
To activate the probe, select the appropriate probe from the probe indicators on the touch panel.
The probe's default settings for the mode and selected exam are used automatically.
3-7-4-5
Deactivating the Probe
When deactivating the probe, the probe is automatically placed in standby mode.
To deactivate a probe:
1.) Ensure the LOGIQ S7 Expert/Pro is in freeze mode. If necessary, press the
Freeze
key.
2.) Gently wipe the excess gel from the face of the probe.
3.) Ensure that the probe is placed gently in the probe holder.
3-7-4-6
Disconnecting the Probe
Probes can be disconnected at any time. However, the probe should not be active when disconnecting
the probe.
1.) Ensure the probe is deactivated. Deactivate by selecting another probe or pressing Freeze.
2.) Move the probe locking handle to the left.
3.) Pull the probe connector straight out of the probe port carefully.
4.) Ensure the cable is free.
5.) Be sure that the probe head is clean before placing the probe in its storage box.
CAUTION
!! CAUTION:
Make sure that the probe and application names displayed on the screen correspond to the
actual probe and application selection.
CAUTION
!! CAUTION:
DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable
damage. Use the integrated cable management hook to wrap the cord.