8-134
Section 8-32 - Replacement of Filter
8-31-4
Installation Procedure
In reverse order of removal.
8-31-5
Functional Check
1.) Visually check LCD monitor.
2.) LCD Arm Test. (Refer to
Section 4-3-2-2
).
Section 8-32
Replacement of Filter
8-32-1
Manpower
1 person, 2 minutes.
8-32-2
Tools
None Required.
8-32-3
Removal Procedure
1.) Remove Front Cover.
2.) Pull out Filter upward.
8-32-4
Installation Procedure
In Reverse Order.
8-32-5
Functional Check
Visual inspection Only.
Section 8-33
Replacement of SW CD set
8-33-1
Manpower
1 person, 3 minutes.
8-33-2
Tools
Phillips Screw Driver.
8-33-3
Removal Procedure
1.) Remove Rear Cover.
2.) CD is placed on Rear Cover.
8-33-4
Installation Procedure
Reverse Order.
8-33-5
Functional Check
Visual inspection Only.