Chapter 4 - Functional Checks
4-16
4-3-3-5
Power Off / Shutdown Check
1.) When you shut down the system, enter the scan screen and lightly press the Power On/Off switch
(see:
Figure 4-14
) once. The System-Exit window is displayed.
2.) The SYSTEM - EXIT menu, used when power off the unit, gives you these choices:
•
Logoff
Use this button to log off the current user. The system remains ON and ready for a new user to log
on. If the Logoff button is dimmed, it indicates that no user is logged on to the unit at the moment.
•
Shutdown
Use this button to shut down the system. The entire system will shut down. It is recommended to
perform a full shutdown at least once a week.
If the Shutdown button is dimmed, press the power button or Alt F10 to shut down the unit.
•
Cancel
Use this button to exit from the System-Exit menu and return to the previous operation.
•
Exit
(Only available when logged in as GE Service with Service Dongle)
Select this button when you want to exit to the Windows Desktop.
Summary of Contents for LOGIQ S7 Expert
Page 2: ......
Page 11: ...GE HEALTHCARE LOGIQ S7 EXPERT PRO DIRECTION 5460683 REVISION 3 SERVICE MANUAL ix ZH CN KO ...
Page 434: ......