Section 8-15 - Replacement of Key Caps (by special native language keys)
8-15-5
Replacement Checking Procedure
1.) Insert service dongle key in the front USB port of OPIO.
2.) Push the power button and then click the exit.
3.) Exit to windows mode by inputting the password.
4.) Run the OPIO Test tool (ScFpApiTest.exe) in the desktop screen and then click the “Op-panel
Open” button of test tool. (Program path: C:\scanner\target\bin\ScFpApiTest.exe)
Figure 8-65
5.) Check on the “Keyboard Test on” check box.
6.) Check that pressed key is displayed except for ‘Fn’ key, when AN Keyboard is pressed.
7.) End ScFpApiTest by click the ‘Op-Panel Close’ button.
Figure 8-66