Section 8-26 - Replacement of Peripheral Cable Assy
3.) Remove the printer USB and power cable connectors from the printer.
Figure 8-129 Remove printer USB and power cable connectors from printer
8-26-2-4
Installation Procedure
1.) Parts to be installed in reverse order of removal.
8-26-2-5
Functional Check
Test 1: Print an image by pressing the button P1 on the operator panel. (P1 key should be configured
correctly)
Verify1: - Verify that image can be printed.
Verify2: - Verify that printed image has no distortion as compared the image on screen.