Chapter 4 - Functional Checks
4-29
3.) Select “Op-Panel Open”.
4.) Verify OP Panel button operation is reflected on diag window.
NOTE:
In above example, key “Alt” is pressed.
NOTE:
Alternatively, OP Panel Interface test may be executed from Common Service Desktop.
However, performance / response of the diagnostics may be degraded when executed with scanner
software still active.
It is located under Diagnostics -> OP Panel Interface in the Diagnostics page.
Before OP Panel Interface test execute from Common Service Desktop, you have to enable the
Disruptive Mode in Diagnostics -> Common diagnostics -> Utilities -> Disruptive Mode
To perform OPIO Interface Check using Diagnostics
1.) Enter Common Service Desktop
Figure 4-32 Op Panel interface test screen