Section 7-4 - Screen Captures and Logs
7-4-3-3
System voltage failure
7-4-3-3-1
Cause
Hardware and/or Voltage error detected.
7-4-3-3-2
What to Do
1.Press OK and reboot the system.
2.If problem persists, then shut down the system, turn OFF breaker and reboot.
7-4-3-3-3
Log message of the Voltage error
Example
Error ; acqHouseKeeping(540); BF_AVDD_3V3_2 = 2.387 outside tolerance limits. UpperBound =
4.000 LowerBound = 2.970, Soft Limits are usually narrower than upperBound and lower bound
Summary of Contents for LOGIQ S7 Expert