8-140
Section 8-36 - Replacement of UVC
Section 8-36
Replacement of UVC
8-36-1
UVC ASSY
8-36-1-1
Manpower
1 person, 1 hour
8-36-1-2
Tools
Standard Field Service Tool kit
8-36-1-3
Removal Procedure
1.) Remove HDMI CABLE form console.
2.) Remove rear cover ASSY with 5 screws and remove cable cover BRKT with 3 screws.
3.) Remove the Y-Cable on the rear panel through the cable cover BRKT hole.