Chapter 2 - Site Preparation
2-3
2-2-2-1
LOGIQ™ S7 Expert/Pro Power Requirements
Power Consumption nominal 900 VA including all on-board peripherals.
2-2-2-2
Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
2-2-2-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
2-2-2-4
Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other adequate
outlets for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-5
Main Power Plug
The LOGIQ™ S7 ultrasound system is supplied with a main power plug, as standard. In the event that
the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary,
the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
Table 2-3
Electrical Specifications for LOGIQ™ S7 Expert/Pro
Voltage
Tolerances
Current
Frequency
100 - 120 VAC
±10%
9.0 ... 7.0 A
50, 60 Hz (±2%)
220 - 240 VAC
±10%
4.0 ... 3.7 A
50, 60 Hz (±2%)
CAUTION
!! CAUTION:
POWER OUTAGE MAY OCCUR.
The LOGIQ™ S7 Expert/Pro requires a dedicated single branch circuit. To avoid circuit overload
and possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.