Section 3-3 - Receiving and Unpacking the Equipment
Section 3-3
Receiving and Unpacking the Equipment
The LOGIQ™ S7 Expert/Pro ultrasound system, together with peripherals, probes and accessories are
shipped from the factory in a single durable shipping crate which is mounted on a raised wooden
platform base.
* Weight is approximate and will vary depending upon the supplied peripherals
Before unpacking the unit:
•
Inspect the crate for visible damage.
•
Inspect the Drop and Tilt Indicator for evidence of accidental shock or tilting during transit (damage
incident, see:
Figure 3-3
below).
CAUTION
!! CAUTION:
Please read this section carefully before unpacking the LOGIQ™ S7 ultrasound system and its
(optional) peripherals.
CAUTION
!! CAUTION:
Transport only with forklift or stacker truck.
During transport pay attention to the point of gravity (“tilt and drop” indicator)!
Have two people available to unpack the LOGIQ™ S7 Expert/Pro.
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
result in personal injury, and/or damage to the system.
Table 3-4
Shipping Carton - Dimensions and Weight
Description
Height
Width
Depth
Weight
*
LOGIQ™ S7 Expert/Pro
incl. peripherals and accessories
1490 mm / 58.6 inch 780 mm / 30.7 inch 1180 mm / 46.5 inch
196 kg / 432 lbs
Figure 3-3 Drop and Tilt Indicator
NOTICE
!! NOTICE:
The device must only be transported in the original packaging!
Each shipping crate is composed of wooden base, cardboard box, re-usable plastic caps and cushions
inside. It is recommended to keep and store the shipping crate and all other packing materials (including
the support foams, anti-static plastic cover, etc.), in case the unit has to be moved to a different location.
Unpack the devices such a way that packaging can be reused.
For warranty purposes, storage of the above is required for one year from date of purchase.
NOTICE
!! NOTICE:
If the shipping crate is damaged, please inform the GE Healthcare sales representative immediately.
Indicator turns