Chapter 8 - Replacement Procedures
8-57
4.) Remove a connector from location [A].
5.) Remove a connector from location [B].
6.) Remove a connector from location [C].
Figure 8-47 Removal of cable & connectors
7.) OP Panel can now be separated from the console.
8-13-3-3
Removal Procedure - AN Keyboard
1.) Remove connector from back of the AN Keyboard (marked below).
2.) Remove 6 screws around AN Keyboard.
3.) Slowly lift up AN Keyboard from OP Panel
.
Figure 8-48 AN Keyboard
Location
Connector
Description
A
CN1
Power/Signal to Host
B
CN2
Signal to ECG
C
J2
Power to Gel warmer